WHICH SECTION ARE YOU INTERESTED IN?
HOW TO USE IT
▶️ Connect one end of the tube to the air inlet.
▶️ Open the top part of the nebulizer by rotating it counterclockwise, then lift it.
▶️ Pour the prescribed medication into the bottom of the nebulizer. The maximum fill volume is 8 ml.
▶️ Close the nebulizer lid. The semicircle should be facing forward.
▶️ Attach the mouthpiece or mask to the nebulizer outlet.
▶️ Connect the free end of the air tube to the bottom of the container.
▶️ Turn on the device when you are ready to start the treatment.
▶️ For use with the mouthpiece, close your lips around the mouthpiece and inhale slowly.
▶️ For the mask, position the mask over your nose and mouth and inhale slowly.
CLEANING AND MAINTENANCE
CLEANING THE COMPRESSOR
▶️ Use a clean, dry cloth and mild detergents to clean the exterior of the unit.
⚠️ WARNING ⚠️
▶️ During cleaning, ensure that the internal parts of the unit do not come into contact with liquids and that the power cord is not plugged in.
▶️ Never open the device: there are no repairable parts inside the unit. The compressor does not require lubrication or maintenance.
HOW TO DISINFECT AND CLEAN ACCESSORIES
• The medication container can be washed with water and liquid soap and then disinfected by boiling it in water (max. 10 minutes).
• It is recommended to disinfect the mask and mouthpiece using cold disinfectant liquids.
• Dry the nebulizer using a clean cloth.
• After drying, connect it to the air outlet and turn on the compressor to check that it functions after cleaning.
❌ WHAT NOT TO DO
• Do not boil the mouthpiece, air tube, or masks.
NEBULISER REPLACEMENT
With normal use, the container should last at least 6 months.
FILTER REPLACEMENT
Open the filter holder to replace the worn filter with a new one.
Use only original filters and above all, never use the device without a filter.
FREQUENTLY ASKED QUESTIONS
The device does not switch on
• Check that the plug is properly inserted into the electrical outlet.
• Ensure that the device has operated within the operational limits specified in the instruction manual, namely 30 minutes ON and 30 minutes OFF.
The device does not atomise or the atomisation rate is weak
• Ensure that the ends of the air tube are securely attached to the main unit and the nebulizer.
• Check if the medication container is empty or overfilled.
• Check if the nebulizer nozzle is clogged.
• Verify that the deflector is inserted correctly.