PORTABLE PISTON AEROSOL 5030


INFO CENTER

HOW TO ASSEMBLE THE AEROSOL


WHAT TO DO

▶️ Lift the mouthpiece, nosepiece or mask from the medicine container and unscrew the top shell of the nebulizer.
▶️ Insert the correct amount of the prescribed medication into the nebulizer container.
▶️ Screw the top nebulizer shell onto the bottom nebulizer shell clockwise until it is securely closed.
▶️ Install the desired inhalation accessories.
▶️ Rotate the air hose plug and Push it firmly into the air hose connector of the compressor.
▶️ Rotate the air tube plug slightly and push it firmly into the air tube connector on the bottom of the medicine container.

⚠️ ATTENTION ⚠️

Keep the medicine container in place vertical when assembling the air hose, otherwise the medicine may leak out of the medicine container.

INDICATOR LIGHTS

State of charge

• Red light 🔴 on: charging in progress.
• Red light 🔴 off: charging complete.
• Red 🔴 and blue 🔵 light on at the same time: the device is charging while It's running.

During use

• Blue light 🔵: The device is working.
• Blue light 🔵 flashing: low battery, the device is about to shut down automatically, please charge it as soon as possible.

HOW TO USE



▶️ Hold the medicine container as shown in the figure in the manual.
▶️ Turn on the unit by pressing the power button for 0.5 seconds. When the compressor starts, the nebulization begins.
▶️ Inhale the medicine following the doctor's instructions.
▶️ Turn off the unit by pressing the power button for 0.5 seconds. When the compressor stops, the nebulization will also stop.
▶️ Disconnect the air tube from the medicine container.
▶️ Disconnect the air hose from the compressor.

HOW TO USE THE ACCESSORIES (MOUTHPIECE, MASK, NOSE PIECE)

Mouthpiece

• Insert the mouthpiece into your mouth.
• Breathe in normally the drug.
• Exhale normally through the mouthpiece.

Mask

• Place the mask over your nose and mouth.
• Pull the elastic strap over your head.
• Pull gently the strap to secure the mask to the nose and mouth.
• Inhale the medicine.
• Exhale normally through the mask.

Hake

• Inhale the medicine through the nosepiece as illustrated in the manual.
• Exhale through your mouth.

HOW TO DISCONNECT THE AIR TUBE

• Hold the air tube plug and gently pull it down.
• Check the air hose, making sure there is no condensation or moisture left inside.

If any condensation or moisture remains in the air pipe, remove it following the instructions given below:

• Make sure the air hose is still connected to the compressor air hose connector.
• Turn on the unit and the compressor will start and pump air through the air hose to push out the moisture.
• Turn off the unit.

CLEANING



▶️ Remove the inhalation accessory from the medicine container.
▶️Disconnect the air tube from the medicine container.
▶️ Gently unscrew the top shell of the medicine container.
▶️ Discard any remaining medication.
▶️ Wash the parts in water.
▶️ Hand dry or air dry in a clean environment using a soft cloth and clean.
▶️ Reassemble the medicine container and store it in a dry bag.

⚠️ ATTENTION ⚠️

▶️ Do not use abrasive cleaners to clean the main unit case. Use a soft cloth moistened with water or mild detergent.

HOW TO DISINFECT IT

✅ WHAT TO DO

• Soak the parts in the disinfectant solution for the specified period.
• Remove parts and discard solution.
• Rinse the parts with clean, warm tap water, shake off excess water and dry them in a clean environment.

❌ WHAT NOT TO DO

• Never dry parts in the microwave.
• Never use an autoclave, EOG gas disinfection or sterilizer low temperature plasma to disinfect the device.

MAINTENANCE



▶️ Do not drop the device or subject it to other shocks or vibrations.
▶️ Never dry parts in the microwave.
▶️ Do not carry or leave the nebulizer with the medicine in the container for medicines.
▶️ Discard any remaining medication.
▶️ Do not disinfect by boiling any parts other than masks and mouthpieces.

FREQUENTLY ASKED QUESTIONS

No power to the device when the switch button is pressed

If the device does not turn on when you press the switch button, the battery is low.
Charge the battery to use the aerosol.

The aerosol does not nebulize or the nebulization speed is low


If the aerosol does not nebulize or the nebulization speed is low, there could be several causes:

There is no medication or there is too much medicine in the container for medicines.
Add quantity indicated by the drug prescribed to solve the problem.

The container for medicine is not assembled correctly.
Make sure that the container for medicines are assembled correctly and that the accessory for inhalation is connected correctly to solve the problem.

The nozzle is blocked.
Clean and disinfect the medicine container to remove the blockage.

The container for medicines is inclined with an angle incorrect.
Keep the container for medicines correctly. Don't tilt it with a angle greater than 45 degrees.

The air tube is incorrectly connected.
Make sure the tube of the air is connected correctly to the compressor and al medicine container.

The air hose is kinked, damaged or blocked.
Make sure the tube of the air is not bent and inspect it for any damage. Replace the tube of the air if damaged to solve the problem.

The air filter is dirty.
Replace the air filter with a new air filter clean to solve the problem.

The device is very hot

If the device is very hot, you may have accidentally covered the compressor. Do not cover the compressor during use to use it properly.

Also, if you use the device for more than 30 minutes it may overheat excessively.
To avoid this problem, we suggest you limit usage to 30 minutes at a time and wait a 20-minute interval before using the device again.

The device is unusually loud

If the device is unusually loud, you may have connected the air hose incorrectly.
To solve the problem, fix the air filter cover correctly, making sure that it is not blocked.