WHICH SECTION ARE YOU INTERESTED IN?
HOW TO USE
▶️ The adult mask has the concave end of the nasal bridge facing upward and the medication chamber of the main body facing upward.
▶️ Grip both sides of the mask to deform it while simultaneously attaching the main unit to the front of the mask to form a single unit and ensure that the buckle is in position.
▶️ The main unit is used with the mouthpiece or the children's mask for portable use.
▶️ For children's masks, install the end with the dented nasal bridge vertically upward. Please keep the nozzle flat in the mouth and install it horizontally.
⚠️ WARNING ⚠️
▶️ The device will automatically turn off after 20 minutes of continuous operation.
▶️ Do not shake the device during use and avoid moving, as abrupt movements may interrupt operation.
▶️ If treatment for the throat is needed, use a mouthpiece; if treatment for both the nasal cavity and throat is needed, choose a mask. The nozzle is suitable for adults and children over 5 years old.
HOW TO WEAR IT
• Hold the body of the mask in one hand; hold the two straps with the other hand and pull.
• Place the mask over the nose and mouth, stretch the two straps, and place them at the back of the neck.
• Pull the outer strap to the top of the head and adjust the position of the two straps to achieve a stable fit.
• If it is too loose or too tight, adjust the strap length and wear it again. If the two loops are located on the upper and lower sides of the ears, the mask is positioned correctly.
HOW TO HOLD IT DURING TREATMENT
• During treatment, it is recommended to keep the device in a vertical position or slightly tilted backward, but not tilted sideways; otherwise, nebulization may not occur properly, leading to the device stopping. In case of interruption, turn the device back on.
HOW TO ADD MEDICATION
• Open the lid of the container upward and pour in the medication.
• Close the lid and secure it with the designated latch.
⚠️ WARNING ⚠️
• Before adding medication from the medication container lid, remove the main unit from the adult mask and operate it.
• After adding the medication, check for bubbles. Bubbles may cause the device to function improperly and stop nebulizing. If bubbles are present, shake the device to remove them.
• If there is a noticeable liquid leak, check if the safety closure of the medication container is secured and if the sealing ring inside the lid of the medication container is loose or missing.
HOW TO USE THE DEVICE WITH CHILDREN
To use the device with young children, we recommend using the silicone support for handheld use and following these instructions:
• Insert the device into the silicone support.
• Attach the children's mask to the device.
• Inhale while holding the support with the device inside, ensuring that it remains in a vertical position at all times.
RECOMMENDATIONS FOR PROPER USE DURING TREATMENT
• When using the Mouthpiece, allow the flat end of the Mouthpiece to enter the mouth with the lips slightly closed and inhale through the mouth.
• When using the mask, it should fit the face and cover both the mouth and nose at the same time.
• Use the mouthpiece correctly, avoiding directing the nebulization onto the tongue. Do not bite the mouthpiece.
• During nebulization treatment, maintain a slow and deep breathing rhythm to inhale more medication.
• During treatment, the medication liquid will condense on the skin. Once the treatment is complete, clean the face promptly.
• Do not swallow the nebulized liquid and keep it as much as possible on the affected area.
LIGHT INDICATOR EXPLANATION
• Blue light 🔵 remains on: the aerosol can be used.
• Red light 🔴 flashing: cleaning function in progress.
• Red light 🔴 blinks during use: low battery.
• Red light 🔴 blinks during charging: battery charging in progress.
• Red light 🔴 steady: charging is complete, disconnect the charging cable.
CLEANING
✅ WHAT TO DO
▶️ Use clean and sterilized medical gauze and cotton swabs.
▶️ If the metal mesh plate is blocked, pour 2ml of warm water (max 60°C) into the medication container, press the power button for 5 seconds to access the cleaning mode, and nebulize until the liquid is exhausted.
▶️ Turn off the device, open the lid of the medication container, and soak it in warm water or a solution of water and vinegar for about 3 minutes, then rinse with pure water.
▶️ Always turn off the power and unplug the power cord during cleaning and disinfection.
▶️ Always open the lid of the medication container when soaking it in water or alcohol for cleaning.
❌ WHAT NOT TO DO
▶️ Do not use paper towels or other cloths to dry the water, as dust or fabric particles may remain on the components.
▶️ Do not touch the mesh with sharp and/or hard objects.
▶️ Do not use cleaning agents other than those recommended.
HOW TO CLEAN AND DISINFECT THE MEDICATION CONTAINER
• Press down the unlock button and simultaneously pull back the container to detach it from the main unit.
• Open the lid of the container after each nebulization and empty it.
• Pour a small amount of warm water into the container (water temperature 40ºC / 50ºC), close the lid, shake it, and empty it.
• Again, pour 2 ml of warm water, hold the power button for 5 seconds to access the cleaning mode.
• Once the warm water is exhausted, press the power button to turn off and clean the remaining medication with a clean medical gauze.
• Rinse the inner and outer walls of the container and the lid with pure water. After cleaning, use a clean medical gauze to gently dry off the water. The vibrating mesh can be cleaned with a medical cotton ball, and the action should be as gentle as possible.
• The medication container can be immersed in 75% medical alcohol for at least 3 minutes for disinfection, then allow it to drain and place it in a clean area to dry.
HOW TO CLEAN THE MAIN UNIT
• If there are stains on the exterior surface of the main unit, dampen a gauze with 75% medical alcohol, gently clean it, then dry it with a dry gauze.
• A dry gauze can be used to clean the electrodes on the main unit and the container to ensure that the electrodes are connected and that the device can function normally.
• After cleaning and drying it, push the medication container forward from the back of the main unit, then push it until it locks onto the main unit.
⚠️ WARNING ⚠️
Do not use other volatile liquids to clean the main body of the device.
The main unit is not waterproof and cannot be washed or immersed in any liquid.
HOW TO CLEAN AND DISINFECT THE MOUTHPIECE AND MASK
• Remove the mouthpiece or the children's mask and the adult mask.
• Rinse with pure water for 2/3 minutes.
• Use 75% medical alcohol to clean all surfaces of the mask and mouthpiece for more than 3 minutes for disinfection.
• The mask and mouthpiece can be sterilized by boiling in hot water.
• Dry off the water with a medical gauze.
• The soft rubber ring of the mask can be disassembled and reinstalled after cleaning.
MAINTENANCE
▶️ Use the device under normal working conditions and only for personal use.
▶️ Do not use the device near a heating device or open flames.
▶️ Do not dry the device and its accessories in the microwave, oven, with a hairdryer, etc...
▶️ Avoid contact between the nebulizer and its accessories with corrosive liquids or gases.
▶️ Do not place the main unit near power lines.
▶️ Clean and disinfect the device after each use.